PY823 intelligent pressure controller uses digital control technology, intelligent interaction technology, rapid output of high precision and high stability of pressure, meet the industrial field and laboratory calibration or test pressure gauge, common and intelligent pressure transmitter, pressure switch requirements. It fully conforms to the metrological verification regulation of JJG1107-2015 automatic standard pressure generator.
Product features
■ Built-in positive pressure source and negative pressure source, automatic output set pressure;
■ Voltage, millivolt, current, switch (on and off) and other electrical measurement;
■ Built-in DC24V power supply, verification transmitter without external power supply;
■ Built-in HART communication, can be calibrated HART instrument;
■ Built-in standard pressure module, optional external pressure module;
■ 7 inch touch screen, Chinese and English display, good human-computer interaction interface;
Technical parameters
Display: 7-inch TFT touch screen
Interaction: Support color screen touch and keyboard operation
Power supply: DC24V adapter
Pressure connection: M20*1.5 quick connection internal thread specifications, different versions
Electrical connection: ø 4 banana jack
Communication: USB, WIFI and Bluetooth
Storage: 8G
Task management: Can manage more than 1000 tasks
Unit of pressure: 21 common units
Operating system: WINCE
Size: 420 * 275 * 240 mm
Weight: 12 kg
Use environment:
Ambient temperature: (-20~70)℃
Temperature compensation: (0~50)℃
Relative humidity: <95%
Atmospheric pressure: (86~106)kPa
Storage temperature: (-40~85) ° C
Pressure indicators:
version | PY823-100K | PY823-600K |
Pressure occurs in the inverse range1 | (-100~100)kPa | (-100~600)kPa |
Pressure control module I | 0.02级 | 0.02级 |
Pressure control module II | One or more pressure modules can be purchased. The range of pressure modules cannot exceed the range of built-in pressure modules |
Pressure type | Differential pressure, gauge pressure |
Atmospheric pressure module accuracy | ±55Pa |
Control stability 2 | ≤0.003%F.S |
Control response time ³ | <30S |
[1] when the ambient atmospheric pressure is 100kPa.A;
[2] The control stability is based on the range of pressure control module;
[3] When no load volume is applied and 15% step of the range is completed.
Electrical measurement index (ambient temperature 20℃±5℃, 1 year accuracy) :
Project | Range | resolution | Indicators |
measurement | Voltage measurement | (-500~500)mV | 1uV | ±(0.008%RD+6uV) |
Voltage measurement | (0~30)V | 0.1mV | ±(0.008%RD+600uV) |
Current measurement | (0~30)mA | 0.1uA | ±(0.008%RD+1uA) |
Switch measure |
| N/A | Voltage range if switch is live(1~12)V |
Bus communication | Support HART bus communication |
KZDM pressure module selection table:
Ordering code | Pressure range |
Accuracy | medium |
KZ2.5K | ±2.5kPa | 0.02 | gas |
KZ 10K | ±10 kPa | 0.02 | gas |
KZ 16K | ±16 kPa | 0.02 | gas |
KZ 25K | ±25 kPa | 0.02 | gas |
KZ 40K | ±40 kPa | 0.02 | gas |
KZ 60K | ±60 kPa | 0.02 | gas |
KZ 100K | (-100~100) kPa | 0.02 | gas |
KZ160K | (-100~160) kPa | 0.02 | gas |
KZ250K | (-100~250) kPa | 0.02 | gas |
KZ400K | (-100~400) kPa | 0.02 | gas |
KZ600K | (-100~600) kPa | 0.02 | gas |